Evaluation of the implant diameter on the initial-stability of narrow- and standard-diameter implants placed in simulated Type-I and Type-IV bone-blocks

  • Saad Alresayes King Saud University
  • Sameer A Mokeem King Saud University
  • Aasem M Alhenaki King Saud University
  • Fahim Vohra KING SAUD UNIVERSITY
  • Tariq Abduljabbar King Saud University
Keywords: Dental implant, in-Vitro, initial stability, Simulated bone, narrow diameter, Standard diameter

Abstract

Objective: A comparison of the initial stability of narrow- and standard-diameter implants (SDIs) placed in Type-I and Type-IV bone-blocks is not yet reported. The aim was to evaluate in-vitro the influence of implant diameter on the initial stability of narrow- and standard-diameter implants (SDIs) placed in simulated Type-I and Type-IV bone-blocks.

Methods: The present experimental in-vitro study was performed between July and September 2020 at the Specialist Dental Practice, Riyadh, Saudi Arabia. Narrow- and standard-diameter implants were placed 3-mm apart in simulated soft (Type-IV) and dense (Type-I) bone blocks by a trained and calibrated investigator. In groups A (Type-IV bone blocks) and B (Type-I bone blocks), implants were inserted using an insertion-torque and drilling-speed of 15-30 Ncm and 1000-1500 rpm, respectively with the implant collar at the crest of simulated bone blocks. In all samples, initial-stability was recorded using resonance frequency analysis (RFA). Sample-size estimation was done and group-comparisons were carried out. A P-value of 0.01 or less reflected statistical significance.

Results: In Groups-A and -B, 44 (22 NDIs and 22 SDIs) and 44 (22 NDIs and 22 SDIs) were placed. In group-A, the mean RFA values for NDIs and SDIs were 68.5 ± 3.5 and 69.1 ± 2.4, respectively. In Group-B, the mean RFA values for NDIs and SDIs were 78.06 ± 9.6 and 75.3 ± 5.2. RFA values among NDIs and SDIs in groups A and B were similar.

Conclusion: The NDIs and SDIs show comparable initial-stability when positioned in simulated Type-I and Type-IV bone blocks.

doi: https://doi.org/10.12669/pjms.37.3.3943

How to cite this:
Alresayes S, Mokeem SA, Alhenaki AM, Vohra F, Abduljabbar T. Evaluation of the implant diameter on the initial-stability of narrow- and standard-diameter implants placed in simulated Type-I and Type-IV bone-blocks. Pak J Med Sci. 2021;37(3):812-815.  doi: https://doi.org/10.12669/pjms.37.3.3943

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Published
2021-03-26
How to Cite
Alresayes, S., Mokeem, S. A., Alhenaki, A. M., Vohra, F., & Abduljabbar, T. (2021). Evaluation of the implant diameter on the initial-stability of narrow- and standard-diameter implants placed in simulated Type-I and Type-IV bone-blocks. Pakistan Journal of Medical Sciences, 37(3). https://doi.org/10.12669/pjms.37.3.3943
Section
Original Articles

Most read articles by the same author(s)